| Dimensions | : 225 mm (W). 225 mm (H). 69 mm (D). | 
| Weight | : 1.2 Kg (basic) to 1,9 Kg (complete). | 
| Non-rechargeable internal battery | : Type: Manganese Lithium Dioxide (LiMnO2) 18V, 1400 mAh (optional 1) or 2800 mAh (optional 2). | 
|  | : Duration of optional battery 1 (1400 mAh): more than 80 shocks in 200 J or 6 hours of continuous monitoring. | 
|  | : Duration of optional battery 2 (2800 mAh): more than 300 shocks in 200 J or 15 hours of continuous monitoring. | 
| Rechargeable internal battery | : Type: Li-Ion, 14.4 VDC 4.0 A/h. | 
|  | : Duration: 18 hours in heart rate recognition mode (fully charged battery), or a minimum of 400 shocks in 200 J. | 
|  | : Time for battery full charge (fully discharged): 5 hours. | 
| Battery power supply charger | : Power 100 – 220 V/50-60 Hz. | 
|  | : Consumption (maximum): mains supply 1 A. | 
|  | : Output: 16.8 VDC, 1 A. | 
|  | : Use only source manufactured by Instramed. | 
| Battery storage | : Storing the battery for a long period of time in temperatures higher than 35°C (95°F) will reduce its capacity and shelf life. | 
| Pre-adjusted defibrillation scales | : Adult (automatic): 1º shock – 150 J; following shocks – 200 J. | 
|  | : Adult (manual): up to 360 J. | 
|  | : Child: 50 J. | 
|  | : Defibrillation scales adjusted by user (via SoftDEA): | 
|  | : Adult (non-rechargeable battery 1400 mAh): | 
|  | : Adult (non rechargeable battery 2800 mAh or rechargeable): scales between 120 J and 360 J. | 
| Internal memory storage | : 100 events or 2 hours of ECG recording. | 
| Ambient sound storage | : Up to 10 hours (optional). | 
| Protection rating | : IP56. | 
| Classification | : Class II, internally energized equipment. | 
| Electrical insulation | : CF type. | 
| Operating mode | : Continuous operation. | 
| Maximum time from rhythm analysis… | : 200 J: 20 seconds. | 
|  | : 360 J: 25 seconds. | 
| Maximum time from rhythm analysis… | : 200 J: 25 seconds. | 
|  | : 360 J: 35 seconds. | 
| Insulation method from main power supply | : Flexible cable with a mains plug. | 
| Non-frequent use equipment | : Comply with the requirements for non-frequent use equipment as specified in NBR IEC 60601-2-4 standard. | 
| ENVIRONMENTAL SPECIFICATIONS |  | 
| Temperature | : Operational: 0 to 50ºC. | 
|  | : Storage: -20 to 50ºC. | 
| Humidity | : Operational: 10 to 95% RH, without condensation. | 
|  | : Storage: 10 to 95% RH, without condensation. | 
| Altitude | : Recommended to use in altitudes lower than 2000 meters. | 
| DEFIBRILLATOR |  | 
| Waveform | : Biphasic truncated exponential. Wave shaped parameters adjusted according to the patient’s impedance. | 
| Shock application | : By means of multifunctional adhesive pads. | 
| Commands | : Front panel button: on/off. | 
|  | : Touchscreen: allows for the selection of manual mode and defining energy scales. | 
| Scales for defibrillation | : Adult: 150 to 360 J. | 
|  | : Child: 10, 20, 30, 40 or 50 J. | 
| Adults/children selection | : Automatic due to the size of the pads. | 
| Charge command | : Automatic after identifying arrhythmias that should receive shock. | 
| Shock command | : Front panel button, when blinking. | 
| Maximum time from rhythm analysis… | : 200 J: 20 seconds. | 
|  | : 360 J: 25 seconds. | 
| Maximum charging time (rechargeable…) | : Rechargeable battery… | 
 
There are no reviews yet.