Dimensions |
: 225 mm (W). 225 mm (H). 69 mm (D). |
Weight |
: 1.2 Kg (basic) to 1,9 Kg (complete). |
Non-rechargeable internal battery |
: Type: Manganese Lithium Dioxide (LiMnO2) 18V, 1400 mAh (optional 1) or 2800 mAh (optional 2). |
|
: Duration of optional battery 1 (1400 mAh): more than 80 shocks in 200 J or 6 hours of continuous monitoring. |
|
: Duration of optional battery 2 (2800 mAh): more than 300 shocks in 200 J or 15 hours of continuous monitoring. |
Rechargeable internal battery |
: Type: Li-Ion, 14.4 VDC 4.0 A/h. |
|
: Duration: 18 hours in heart rate recognition mode (fully charged battery), or a minimum of 400 shocks in 200 J. |
|
: Time for battery full charge (fully discharged): 5 hours. |
Battery power supply charger |
: Power 100 – 220 V/50-60 Hz. |
|
: Consumption (maximum): mains supply 1 A. |
|
: Output: 16.8 VDC, 1 A. |
|
: Use only source manufactured by Instramed. |
Battery storage |
: Storing the battery for a long period of time in temperatures higher than 35°C (95°F) will reduce its capacity and shelf life. |
Pre-adjusted defibrillation scales |
: Adult (automatic): 1º shock – 150 J; following shocks – 200 J. |
|
: Adult (manual): up to 360 J. |
|
: Child: 50 J. |
|
: Defibrillation scales adjusted by user (via SoftDEA): |
|
: Adult (non-rechargeable battery 1400 mAh): |
|
: Adult (non rechargeable battery 2800 mAh or rechargeable): scales between 120 J and 360 J. |
Internal memory storage |
: 100 events or 2 hours of ECG recording. |
Ambient sound storage |
: Up to 10 hours (optional). |
Protection rating |
: IP56. |
Classification |
: Class II, internally energized equipment. |
Electrical insulation |
: CF type. |
Operating mode |
: Continuous operation. |
Maximum time from rhythm analysis… |
: 200 J: 20 seconds. |
|
: 360 J: 25 seconds. |
Maximum time from rhythm analysis… |
: 200 J: 25 seconds. |
|
: 360 J: 35 seconds. |
Insulation method from main power supply |
: Flexible cable with a mains plug. |
Non-frequent use equipment |
: Comply with the requirements for non-frequent use equipment as specified in NBR IEC 60601-2-4 standard. |
ENVIRONMENTAL SPECIFICATIONS |
|
Temperature |
: Operational: 0 to 50ºC. |
|
: Storage: -20 to 50ºC. |
Humidity |
: Operational: 10 to 95% RH, without condensation. |
|
: Storage: 10 to 95% RH, without condensation. |
Altitude |
: Recommended to use in altitudes lower than 2000 meters. |
DEFIBRILLATOR |
|
Waveform |
: Biphasic truncated exponential. Wave shaped parameters adjusted according to the patient’s impedance. |
Shock application |
: By means of multifunctional adhesive pads. |
Commands |
: Front panel button: on/off. |
|
: Touchscreen: allows for the selection of manual mode and defining energy scales. |
Scales for defibrillation |
: Adult: 150 to 360 J. |
|
: Child: 10, 20, 30, 40 or 50 J. |
Adults/children selection |
: Automatic due to the size of the pads. |
Charge command |
: Automatic after identifying arrhythmias that should receive shock. |
Shock command |
: Front panel button, when blinking. |
Maximum time from rhythm analysis… |
: 200 J: 20 seconds. |
|
: 360 J: 25 seconds. |
Maximum charging time (rechargeable…) |
: Rechargeable battery… |
There are no reviews yet.