Energies |
: Standard max 200J or Power max 360J |
Waveform |
: Adaptive BTE (biphasic truncated exponential) conforming to patient chest’s impedance |
Protocols |
: Various adult shock protocols available on request |
Factory default |
: Adult Standard escalating 150, 200, 200J |
|
: Adult Power escalating 200, 250, 360J |
|
: Pediatric (Standard or Power) 50J fixed |
Charging time |
: ≤9 seconds with a new and fully charged battery |
|
: Depleted battery will result in a longer charging time |
Analysis time |
: IEC/EN 60601-2-4 from 4 to 15 seconds |
Impedance |
: 20-200 ohms |
Sensitivity |
: IEC/EN 60601-2-4 (AHADB, MITDB source), 97% |
Specificity |
: IEC/EN 60601-2-4 (AHADB, MITDB source), 99% |
Controls |
: 3 buttons : ON/OFF, Shock , “i“ info button |
Flashing Icons |
: “connects pads to patient“ |
|
: “adult/child“ informing on pads type use |
|
: “touch patient“ informing it’s safe to touch |
Indicators |
: Status LED indicator informing on device condition |
|
: Battery gauge with remaining capacity rate |
|
: Audible alerts and text display with service alarms |
Upgradeable |
: Through a USB cable or memory card |
Battery Options |
|
Type |
: Li-SOCl2 Disposable, code SAV-C0903 |
Autonomy |
: 300 complete rescue cycles (shocks at 200J and CPR) or 200 complete rescue cycles (shocks at 200J and CPR) or 35 hours ECG analysis for a new and fully charged battery (*) |
Shelf-Life |
: When stored in original packaging 5 years (*) |
Battery-Life |
: 4 years once installed to AED, assuming one battery insertion test and daily self-test but without switching AED on (*) |
Type |
: Li-ion Accumulator, code SAV-C0011 |
Recharging time |
: 2,5 hours with the charger station code SAV-C0014 (*) (recommended to charge every 4 months at least) |
Autonomy |
: 250 shocks at 200J or 160 shocks at 360J or 21 hours in ECG analysis for a new fully charged accumulator (*) |
Battery-Life |
: 2 years or 300 charging cycles (*) |
Pads Options |
|
Type |
: Disposable, pre-gelled and self-adhesive |
Adult |
: Code SAV-C0846, for patient >8 years or >25 kg |
Pediatric |
: Code SAV-C0016, for patient <8 years or <25 kg |
Cable length |
: 120 cm |
Shelf-Life |
: 30 months |
Event Recording |
|
Internal memory |
: Up to 6 continuous hours of ECG and rescue events |
Optional memory |
: Removable SD card; length of storage depends on card capacity: a 2GB card records up to 100 hours |
Data recording |
: “AED1LOG“ text file with detailed self-test activity “AEDFILES“ with complete recorded information |
Event review |
: “Saver View Express“ data manager software |
Physical |
|
Size |
: 26,5 x 21,5 x 7,5 cm |
Weight |
: 1,95 kg with disposable battery |
|
: 2,00 kg with rechargeable battery |
Environmental |
|
Operating temperature |
: 0°C to 55°C (32°F TO 131°F) |
Storing/Shipping temperature |
: -40°C to 70°C (-40°F TO 158°F) without battery |
Humidity |
: 10% to 95% relative humidity non condensing |
Sealing (IP Protection) |
: IEC/EN 60529 class IP54; splash proof, dust protected |
Shock/Drop Abuse Endurance |
: IEC/EN 60601-1 clause 21; 1 meter drop, impact, force, rough handling, mobile tolerance |
Electrostatic Discharge |
: IEC/EN 61000-4-2 |
Electromagnetic Compatibility |
: IEC/EN 60601-1-2 Emission, Immunity |
Electrical Protection |
: IEC/EN 60601-1; Internally Powered, Type BF |
Directive |
: 93/42/CEE and 2007/47/CE: Class IIb |
There are no reviews yet.